Blaze Labs Research Menu
Location: Experiments > Experiment 09
Home
Food for Thought
Introduction
The Particle Part I
The Particle Part II
Unified Theory Foundations
The EMRP gravity theory
Does a non-linear electric field gradient generate gravity? Generating X-rays
EHD Thrusters
Introduction
EHD thruster collection
Thrusters performance
Lifters in vacuum
Full mathematical analysis (PDF) Ionocraft patent (HTML) Lifter D.O.E. Autonomous thruster project Autonomous thruster calculator Electromagnetic Kinetic Analyser
New Energy Research
Free Energy & Perpetual motion Project Ixion Aquafuel™ generator Transmutation of carbon New fuel from water & carbon
Experiments
Introduction 01: Inertia device 02: Teflon coated EHD thruster 03: Lightweight hv supply 04: Remote controlled lifter 05: Radiation tests on lifters 06: Ionocrafts vs Lifters 07: Heated cathode lifter 08: Zinc vs Aluminium collector 09: Measuring pulsed dc sources 10: EM Magnus effect 11: Gravity Shielding 12: Thruster gas tests 13: High power hv power supply 14: 100g payload lifter design 15: Cockcroft Walton multiplier 16: 300kV helical resonator 17: 50kV lab power supply 18: Ion Triodes 19: Blazelabs Resonant Multipliers 20: Project RX-7 Instrumentation
Infobase
Blaze Labs Wiki Links Database PDF Library Periodic table
Contact us




Experiment 09 (24/10/02) - Measurement errors with pulsed dc sources

© Engineer Xavier Borg - Blaze Labs Research

The aim of this experiment is to confirm the error between the real voltage and the reading of a DC voltmeter with signals of different waveforms, and most important how to obtain the real measurement. This is important, because when we read the hv voltage that we supply to our hv devices, the hv probe is not the only device which introduces errors. One has to consider whether the DC signal being measured is pure dc or any type of pulsed dc.


Equipment used

Signal generator set at 1 V peak dc @ 20 kHz
Oscilloscope set at 0.1 V/div
Standard (not True RMS type) Multimeter


Setup

The signal generator is set at 1 V peak output, and at a frequency of 20 khz. The output from the signal generator is fed to an oscilloscope in parallel with a multimeter, in parallel with a 10 kOhm resistive load. Various waveforms are then selected to check the results on both oscilloscope and multimeter.


Pulsed sine wave testing

This is the kind of waveform one would normally get from a monitor's hv output at 20 khz. As you can see, at 1 V peak signal (as shown on the oscilloscope), the multimeter shows 0.226 V. If this was a reading from a hv probe, one would read 22 kV, when the actual peaks reach 100 kV! Of course this reading varies according to the multimeter frequency response, and duty cycle of the signal.

pulsed sine signal

Pulsed square wave testing

This time, a square wave of 20 kHz, 1 V peak is fed to both meter and oscilloscope in parallel. As you can see, at 1 V peak signal (as shown on the oscilloscope), the multimeter shows 0.436 V. If this was a reading from a hv probe, one would read 43.6 kV, when the actual peaks reach 100 kV!

pulsed square signal

Pulsed triangular wave testing

This time a triangular dc waveform at 20 kHz, 1 V peak is fed to the instruments. At 1 V peak signal (as shown on the oscilloscope), the multimeter shows 0.193 V. If this was a reading from a hv probe, one would read 19 kV, when the actual peaks reach 100 kV!

pulsed triangular signal

Conclusions

  1. The voltage of a pulsed dc source cannot be simply measured by connecting an hv probe to a multimeter.
  2. Errors will depend on type of multimeter used. Its input capacitance, and its internal integration circuit may give further frequency dependent errors, which get worse with higher frequencies.
  3. The best way to measure pulsed dc, would be to first obtain the 'derating' factor for the particular waveform, by using a similar setup shown above, and then apply the factor to the voltages read from the hv probe. For example, if the voltage of the primary winding of my hv flyback was as shown in the pulsed sine experiment above, then I would apply a derating factor of 1/0.226 to the multimeter reading to get the correct value.
Previous Home Next